Colour & UV Service
COLOUR UV Service
Calibration Experts of Spectrophotometers & Colour Booths
647.838.4491
QUOTE
Colour & UV Service
COLOUR & UV Service
Calibration Experts of Spectrophotometers & Colour Booths
647.838.4491

VeriColor

X-Rite VeriColor

Online Spectrophotometer


In-Line Spectrophotometer

VeriColor Spectro Advantages


High Spectral Resolution. True 31-band spectrophotometer provides absolute color data across the visible spectrum at 10 nm intervals


Industrial Hardened. Lab-grade performance in a robust industrially hardened design, NEMA 4 / IP67 rated, that withstands shock, vibration and thermal variation


Patented Technology. Allows accurate in-line capture of color data in the presence of depth variation and dramatic changes in ambient light


Flexible. System interfacing for PC or PLC based operations


Intuitive. Includes easy to use windows based Set-Up and Monitoring Software with real-time visual monitoring and trending graphics


In-line Monitoring. Monitor, control and log color during production — in line, real time


Analyze and contain color problems — make corrections without stopping production


Minimize scrap cost — detect and correct before generating excessive waste


Consistency. Insure consistent color quality, all the time


Specifications


Warm-Up Time

2 – 3 minutes typical at 23° C (73.4°F)


Measurement Time

750 ms


Cycle Time

1 sec. (time interval between measurements)


Ambient Light Rejection

3000 Lux


Repeatability (white tile)*

0.03 avg, / 0.05 max ΔEab

(20 measurements at 3 sec. intervals on a white ceramic tile)


Long Term Repeatability

0.15 avg, / 0.20max ΔEab (over the calibration interval)


Inter-Instrument Agreement Avg E CIELab*

0.30 avg., 0.5 ΔEab max. based on 12 BCRA Series ll tiles


Spectral Range

400 nm – 700 nm


Wavelength Interval

10 nm


Measurement Range

0% - 150% reflectance

Geometry

0/30° , 30/0°

(results based on 0/30°)


Monochromator

Dual beam, 31 channel


LED Life

>10,000,000 measurements


Calibration Interval

Validation recommended – 30 days / 50,000 measurements (whichever comes first)

More frequent verifications may be required if cleanliness of the system is not maintained.


Dimensions

3” x 6” x 9”


Weight: 2.81kg (6.2lbs)


Quote Request: X-Rite VeriColor Spectrophotomter